Instructions & Maintenance for the Lifepak CR2 Defibrillator

Here you'll find instructions and a video on how to get started and maintain your Lifepak CR2 defibrillator.

Taking care of a defibrillator is easy, but if you have any questions, as a Life in Mind customer, you always have access to free phone support and assistance if you have any questions or need help with something like a battery replacement. Here you’ll find instructions on how to get started with your Lifepak CR2, which is our best-selling defibrillator. If you have questions about any of our other defibrillators, you’re always welcome to give us a call.

Get Started with LifePak CR2

  1. Unpack the LifePak CR2
  2. Check the battery's date of manufacture
  3. Check whether the defibrillator has Wi-Fi or cellular connectivity
  4. Register the serial number
  5. Insert the battery and wait until the self-test is complete
  6. Check that the standby indicator is flashing
  7. Check the speaker
  8. See the setup guide for any wireless connection
  9. Keep the defibrillator in an accessible place

Connect to Wi-Fi

  1. Unpack the LifePak CR2
  2. Check the battery's date of manufacture
  3. Check whether the defibrillator has Wi-Fi or cellular connectivity
  4. Register the serial number
  5. Insert the battery and wait until the self-test is complete
  6. Check that the standby indicator is flashing
  7. Check the speaker
  8. See the setup guide for any wireless connection
  9. Keep the defibrillator in an accessible place

Replacing the Electrodes and Battery

Replacing the electrodes:

  1. Open the defibrillator
  2. Press and hold the On/Off button (the button in the middle)
  3. Remove the old electrodes
  4. Insert the new electrodes. Remember to check their best-by dates and seals!
 

Battery Replacement:

  1. Turn the defibrillator upside down
  2. Remove the battery (located below the warning triangle)
  3. Insert a new battery. Remember to check the date of manufacture.
 

Note! When you receive a new battery, it usually comes with a small black piece that you should attach to the old battery. This is to completely discharge the battery before you recycle it. Be careful not to accidentally attach this to the new battery, as it will discharge it. 

Finally: Turn the defibrillator so that the front is facing up, and wait until the self-test is complete; this takes about 2 minutes. The defibrillator is then ready for use.

How to Use the Lifepak CR2
Rent or Buy the Lifepak CR2
CPR Course with a Defibrillator

Help and Support

The defibrillator should be easily accessible and must always be in working order. We’ll, of course, help you with placement and signage, as well as keeping track of when it’s time for battery replacements and refresher training. 

We're always just a phone call away if you have any questions or need help. 

Heart-Safe Zone

The standard is designed to ensure that organizations are well-prepared for sudden cardiac arrest by providing recommendations for training, placement, information, and maintenance of the defibrillator. We help you meet the requirements for a “heart-safe zone,” and the package includes a framed certificate, stickers, and a digital banner for use on your website or in email signatures, for example. 

BFHS

As part of our quality assurance process, we are, of course, members of BFHS (the Swedish Industry Council for Defibrillators). The Industry Council is a supplier- and brand-independent non-profit association for defibrillators and public education, and works to improve survival rates in cases of cardiac arrest.